Autoclave Testing
includes the (1) physical monitoring of the
Steam sterilizer (see
Mechanical indicator)
observing gauges, digital printouts, recorders,
displays and/or gauges, (2) the use of
external and internal
Chemical Indicators
(CI’s) and (3) the frequent (at least weekly)
use of
Biological Indicators
(BI’s).
See
Proof of Sterilization
page showing detail of the three types of monitoring
that is required to be done regularly, some with
every sterilizer cycle. As per the FDA, “a
biological sterilization process indicator is a
device intended for use by a health care provider to
accompany products being sterilized through a
sterilization procedure and to monitor adequacy of
sterilization. The device consists of a known
number of microorganisms, of known resistance to the
mode of
sterilization, in or on a carrier
and enclosed in a protective package. Subsequent
growth or failure of the microorganisms to grow
under suitable conditions indicates the adequacy of
sterilization.” [21 CFR 880.2800(a)(1)]
§
Autoclave Testing Service, Inc. is
one of the first companies nationwide to offer a
Mail-In sterilizer monitoring
service using biological indicators. Serving the
healthcare industry throughout the United States and
Canada since 1986, we have thousands of satisfied
customers.
§
We have
on-line test
results allowing you to access your results
at your convenience! Your Sterilizer Test Report (STR)
is posted
on-line upon
completion of culturing, offering minimum
turn-around time.
§
We have 24-hour test results for
Steam sterilizers!
§
We have a comprehensive Glossary of
Terms related to the sterile processing area (scroll
below to view the Glossary). This is invaluable
information, assisting you with this important area
of your facility.
ATS, Inc. now stocks sterilization
pouches, sterilizer cleaners, chemical indicator
products, sterilizers (new and used), ultrasonic
cleaners, water distillers, and more! (see
Supplies).
In-Office and Mail-To-Lab Testing!
Verify the sterilization process of all steam
sterilizers which include Statim, Midmark,
Tuttnauer, Lisa, Mocom and Pelton & Crane
sterilizers.
Bowie-Dick Test
Packs should be used daily to evaluate the
performance of air-removal (vacuum) system of
dynamic pre-vacuum type sterilizers. Available
as test cards or preassembled test packs.
Internal CI’s should be used within
each pouch or package, tray, or container. Class 5
Integrators should be used in conjunction with
physical monitoring and biological indicator
testing for verifying the efficacy of a
sterilization system. (See AAMI ST79:2006)
Recordkeeping is a must for all
sterile processing departments. This ensures
accountability and allows operators to trace
errors to their source. Tips on sterilizer
maintenance and technical advice are included with
each book.